Scalar i2000 User's Guide
This graphical report is not generated for libraries
Note
with only one frame.
• If the graphical report shows a failed result for the drive-side or door-
side, make sure that all tape magazines are installed properly on that
side and that the calibration targets are correctly snapped on to the
magazines.
• If the graphical report shows a failed result for the I/E station, make
sure the I/E station and front door are completely shut.
• If running the test again still generates failed results, realign the
middle X-axis rail and check the alignment of the top and bottom X-
axis rails at the location of the failure.
See
figure 24
on page 291.
Chapter 5 Maintaining Your Library