Measuring the Thickness
Measuring the Thickness p.140
As an example, here is an explanation of the procedure for measuring the thickness from the top surface of
glass on which specular reflection occurs as the 1st surface to the 2nd surface (as shown in the diagram on
p.140).
Operating procedure
Steps
Key operation
1
2
Set the material of the target to "MIRROR."
Setting the Material of the Target to Measure p.230
If the target object is a transparent object, such as glass, on which specular reflection occurs, set the material of the
measurement object to "MIRROR" to obtain more accurate measurement result.
3
Set the surface to "SUR.1ST."
3-1
3-2
3-3
3-4
3-5
3-6
4
Set the surface2 to "SUR.2ND."
238
Setting Measurement Items
Display
Description
Press and hold the
FUNC mode.
To save the setting, refer to "7-3 Functions of Operating Keys"
(p.222).
Optional
Press the
Press the
Press the
Press the
Select "SUR1."
Press the
Select the measurement surface.
EDGE1 to EDGE4: 1st surface to 4th surface
PEAK: Peak
The current setting value is displayed on the sub-display.
Press
In this example, select "EDGE1" and press the
key for two seconds or more to enter the
/
key to select MEAS, then press the
/
key to select TASK1, then press the
/
keys to select "ITEM" and press the
/
keys to select "THICK" and press the
/
keys to select "TOP" and press the
/
key to enter editing mode and the sub-display blinks.
key.
key.
key.
key.
key.
key.
ZW-8000/7000/5000
User's Manual