Omron ZW-8000 Series User Manual page 155

Confocal fiber type displacement sensor
Table of Contents

Advertisement

Setting the Median Filter
The intermediate value of multiple sets of data can be output as the measurement result.
Item
Setting item
Medial filter
Filter mode
 Multi View Explore
→ Edit pane
→ Task Settings window : [Filter]
1
Select how many of the latest intermediate
values to output from [Medial filter].
Note
The median filter can also be set by the operating keys on the Sensor Controller.
Setting the Median Filter p.242
ZW-8000/7000/5000
User's Manual
Setting value
Description
OFF
Median filter is not used.
(default value)
3
Outputs the intermediate value of the last three measurements.
9
Outputs the intermediate value of the last nine measurements.
15
Outputs the intermediate value of the last 15 measurements.
31
Outputs the intermediate value of the last 31 measurements.
: [Bank Group] | [(Bank Data Name)] (double click)
: [Task Settings] icon (
)
Setting the Output Conditions
153

Advertisement

Table of Contents

Troubleshooting

loading

This manual is also suitable for:

Zw-7000 seriesZw-5000 series

Table of Contents