Omron ZW-8000 Series User Manual page 169

Confocal fiber type displacement sensor
Table of Contents

Advertisement

Setting a Trigger Delay
Set this to ignore measure values following TIMING input to avoid the influence of bounding or mechanical
vibration when a device is started up.
The delay time (the delay between timing input and the start of sampling) and the sampling time can be set.
TIMING input
Sampling period
Measured value
Item
Setting item
Trigger delay
Trigger delay mode
Delay time
Sampling time
 Multi View Explore
→ Edit pane
→ Task Settings window : [Hold]
1
Set the hold mode and trigger.
Setting the Hold Mode p.160
Setting Triggers p.163
2
Check the check box in [Trigger delay]-
[Trigger delay mode].
ZW-8000/7000/5000
User's Manual
ON
OFF
Delay time
Measured values in this time period
is included in the sampling.
Setting value
OFF (default value)/ON
• Internal synchronous measurement
mode
0.02(default value) to 5000[ms]
• External/PDO synchronous measure-
ment mode
1(default value) to 50000[count]
• Internal synchronous measurement
mode
0.02 to 5000[ms]
100(default value)
• External/PDO synchronous measure-
ment mode
1(default value) to 50000[count]
: [Bank Group] | [(Bank Data Name)] (double click)
: [Task Settings] icon (
Sampling time
Description
Turns the trigger delay ON/OFF.
Sets the period(time) between when turns ON
TIMING input and starts sampling.
Measures the measuring count between when turns
ON TIMING input and starts sampling.
Set the time when perform the sampling.
Set the measuring count when performs sampling.
)
Setting the Output Conditions
167