Analog output terminal block

Used for analog output.
Compatible wire specifications: Solid wire 0.2 to 1.5 mm
Class
Name
Analog
OUT (V)
output
OUT (A)
OUT 0V
Important
• Cut the unnecessary signal lines so that they do not touch other signal lines.
Make the wire length as short as possible. Do not use a wire length of over 30 m.
• Analog output is not possible under the following combination of conditions.
Measurement cycle is less than 40 µs.
EtherCAT output is enabled.
• Make sure to fix the terminal block to prevent be injure when pushing the release button with a screw driver.
ZW-8000/7000/5000
User's Manual
2
Stranded wire 0.2 to 1.5 mm
Stranded wire with rod terminal(no plastic sleeve) 0.25-1.5 mm
Stranded wire with rod terminal(with plastic sleeve) 0.25-0.75 mm
AWG 24-16
Pin processed length 10 (±0.5) mm
Signal name
No.
OUT(V)
1
OUT(A)
2
OUT 0V
3
Description
This outputs the measured value, from -10 V to +10 V as the voltage value.
When measurement not possible: Approx. +10.8 V (default value; can be selected by user)
Alarm: Approx. 10.8 V
This outputs the measured value, from 4 mA to 20 mA as the current value.
When measurement not possible: Approx. +20.8mA (default value; can be selected by
user)
Alarm: Approx. +20.8 mA
0 V for analog voltage output.
2
Signal name
4
NC
5
NC
6
NC
2
2
81
Wiring