Calibrating Sensor Head

Calibrate the Sensor Head by obtaining the dark data in the no-incoming light status.
• When removing and inserting a fiber cable from/to the Sensor Controller (Including the initial connection).
• When extending a fiber cable.
• When "Dark" is shown in the main display even though a measurement object exists in the measurement
range.
The Sensor Head should be calibrated without any object within 20mm of the measurement range or with the
tip of the Sensor Head shielded from light with an object with diffuse reflection.
Correct calibration is not possible with a transparent object, semi-transparent object, or mirror.
• Without any object
within the measuring range
Sensor head
Measuring range
1
Set the operation mode to FUNC mode.
3-2 Switching operation modes p.98
 Multi View Explore
→ Edit pane
2
Select the cable length of the extension fiber cable.
48
Basic Operations for Displaying Measurement Results
• With light shielded
: [Device Group] | [(Sensor Name)] | [System] |
[System Data] (double click)
: [Sensor Settings] icon (
Sensor head
Install with the tip of the Sensor Head
shielded by an object with diffuse reflection,
such as metal, paper, or plastic.
Measuring range
)
ZW-8000/7000/5000
User's Manual