Omron ZW-8000 Series User Manual page 32

Confocal fiber type displacement sensor
Table of Contents

Advertisement

ZW-SP/SPR Series
When installing with accessory (installation plate)
1
Place the accessory (installation plate) between the sensor head and the object you want to fix
the sensor head to.
2
Place the sensor head an appropriate distance from the target to measure, and then fix it by tight-
ening four accessory screws (M2 × 10) inserted into their respective installation holes.
Tightening torque: 0.15 N · m
● ZW-SP Series
This installation method can be used to install sensor heads next to each other.
● ZW-SP Series
Important
For the location of screw holes, check the external dimensions in the appendices.
30
Basic Operations for Displaying Measurement Results
M2 screw × 4
● ZW-SPR Series
M2 screw × 4
● ZW-SPR Series
ZW-8000/7000/5000
User's Manual

Advertisement

Table of Contents

Troubleshooting

loading

This manual is also suitable for:

Zw-7000 seriesZw-5000 series

Table of Contents