Agilent Technologies 8860 Maintaining page 109

Gas chromatograph
Hide thumbs Also See for 8860:
Table of Contents

Advertisement

7
Maintaining the Packed Column Inlet
To Clean the Packed Column Inlet
15 Install the septum and septum nut. See
16 Attach the column.
17 Exit maintenance mode. Select Maintenance > Instrument > Perform Maintenance >
18 Reset the septum and glass liner counters.
19 Perform an inlet Leak & Restriction test and reset the maintenance counters.
20 Restore the analytical method.
Maintaining Your GC
Inlet"
on page 99.
Maintenance Mode > Finished.
"To Change the Septum on the Packed Column
109