To Clean The Septum Seat In The Insert Assembly Of The Split/Splitless Inlet - Agilent Technologies 8860 Maintaining

Gas chromatograph
Hide thumbs Also See for 8860:
Table of Contents

Advertisement

5
Maintaining the Split/Splitless Inlet

To Clean the Septum Seat in the Insert Assembly of the Split/Splitless Inlet

To Clean the Septum Seat in the Insert Assembly of
the Split/Splitless Inlet
1 Gather the following:
2 Manually set the inlet and oven temperature to < 40 °C, and wait for the inlet, oven, and
Be careful! The oven and/or inlet may be hot enough to cause burns. If either is hot, wear
WAR N IN G
heat-resistant gloves to protect your hands.
3 Slide the locking tab forward (counterclockwise). Lift the insert assembly straight up and
4 Remove the septum retainer nut or Merlin cap.
5 Use tweezers to remove the septum or Merlin Microseal from the retainer nut. See
52
Replacement septum, see
page 42.
Wrench, hex for changing septum
0- or 00-grade steel wool (optional)
Tweezers
Compressed, filtered, dry air or nitrogen
Wrench, capillary inlet (optional)
other parts you might come into contact with inside the oven, to cool before continuing.
Alternately, place the GC in maintenance mode: Maintenance > Instrument > Perform
Maintenance > Maintenance Mode > Start Maintenance.
away from the inlet to avoid chipping or breaking the liner.
Change the Septum on the Split/Splitless Inlet"
"Consumables and Parts for the Split/Splitless Inlet"
on page 50.