Agilent Technologies 8860 Maintaining page 60

Gas chromatograph
Hide thumbs Also See for 8860:
Table of Contents

Advertisement

5
Maintaining the Split/Splitless Inlet
To Replace the Filter in the Split Vent Trap for the Split/Splitless Inlet
Figure 8.
6 Remove the old filter cartridge and two O-rings.
7 Verify the new O-rings are seated properly on the new filter cartridge.
8 Install the new filter cartridge then reassemble the trap using your fingers to fully tighten
9 The GC maintenance wizard will perform checks at the appropriate times, including Leak &
10 Select Finished, then select OK to exit the GC maintenance wizard.
11 Install the pneumatics cover.
60
Slide trap assembly to expose the filter
the knurled nut. The nut should thread easily.
If the knurled nut does not thread easily, loosen, reassemble the 2 parts of the trap, and
thread again. Do not force it.
Restriction tests, and will automatically reset the maintenance counters.
Filter and O-ring
Knurled nut