Maintaining the Split/Splitless Inlet

To Replace the Gold Seal on the Split/Splitless Inlet

To Replace the Gold Seal on the Split/Splitless Inlet

1 Gather the following:
2 Launch the GC maintenance wizard: Maintenance > Inlets, select the inlet, then select
Be careful! The oven and/or inlet may be hot enough to cause burns. If either is hot, wear
WAR N IN G
heat-resistant gloves to protect your hands.
3 Remove the inlet liner.
4 Remove the column from the inlet. Cap the open end of the column to prevent
Remove insulation cup
Disconnected, capped column
5 Loosen and remove the reducing nut. Remove the washer and seal inside the reducing nut.
Maintaining Your GC
Replacement gold seal, see
page 42.
Replacement washer
1/4-inch wrench (for column)
1/2-inch wrench
Lint-free gloves
Perform Maintenance > Replace Gold Seal (bottom) > Start Maintenance. The wizard will
walk through the steps needed to replace the liner and O-ring (those same steps are
repeated below). Wait for the GC to become ready.
contamination. Remove the insulation cup around the base of the inlet.
"Consumables and Parts for the Split/Splitless Inlet"
on
57