Maintaining the Split/Splitless Inlet
To Clean the Septum Seat in the Insert Assembly of the Split/Splitless Inlet
6 Scrub the residue from the retainer nut and septum holder with a small piece of rolled-up
7 Use compressed air or nitrogen to blow away the pieces of steel wool and septum.
8 Line up the tab on the bottom of the insert assembly with the slot on the inlet body and
9 Firmly press the new septum or Merlin Microseal into the fitting. See
10 Replace the septum retainer nut or Merlin cap and finger-tighten. See
11 Exit maintenance mode. Select Maintenance > Instrument > Perform Maintenance >
12 Select Maintenance > Inlet > Septum injections, then select Reset.
13 Perform a Leak & Restriction test.
14 Restore the analytical method.
Maintaining Your GC
steel wool and tweezers. Do not do this over the inlet.
push down to connect. Slide the locking tab to the left.
Septum on the Split/Splitless Inlet"
Septum on the Split/Splitless Inlet"
Maintenance Mode > Finished.
on page 50.
on page 50.
"To Change the
"To Change the