6
Maintaining the Purged Packed Inlet

To Condition a Packed Column

To Condition a Packed Column

1 Gather the following:
Do not use hydrogen as the carrier for conditioning! It could vent into the oven and present an
WAR N IN G
explosion hazard.
2 Place GC in maintenance mode: Maintenance > Instrument > Perform Maintenance >
Be careful! The oven, inlet, and/or detector may be hot enough to cause burns. If the oven,
WAR N IN G
inlet, or detector is hot, wear heat-resistant gloves to protect your hands.
Wear clean, lint-free gloves to prevent contamination of parts with dirt and skin oils.
C AU T I O N
3 Install the proper liner in the inlet and attach the column. See
4 If present, remove the packed column adapter from the detector base. (If it is installed to
5 Cap the detector fitting with the no-hole ferrule and column nut, or with a blanking nut.
6 Enter a column flow as recommended by the packing manufacturer or an appropriate flow
7 Raise the oven temperature slowly to the conditioning temperature for the column. The
8 Continue conditioning overnight at the final temperature. Cool the oven to room
9 Attach the column to the detector and maintain established flow. See
10 Perform the inlet Leak & Restriction test and reset the maintenance counters.
11 Exit maintenance mode. Select Maintenance > Instrument > Perform Maintenance >
92
Capillary adapter
Column nut and no-hole ferrule or blanking nut
Two 7/16-inch wrenches
1/4-inch open-end wrench
Lint-free gloves
Maintenance Mode > Start Maintenance. Wait for the GC to become ready.
Column"
on page 85.
the column, it can be conditioned with the column.)
as follows:
20 to 30 mL/min for 2-mm id glass or 1/8-inch od metal columns
50 to 60 mL/min for 4-mm id glass or 1/4-inch od metal columns
conditioning temperature is never higher than the maximum temperature limit for the
column; 30 °C less than the maximum is usually sufficient.
temperature with carrier flow on.
Metal Column"
on page 85.
Maintenance Mode > Finished.
"To Install a Packed Metal
"To Install a Packed
Maintaining Your GC