12 Maintaining the NPD

To Maintain the NPD Collector, Ceramic Insulators, and Jet

To Maintain the NPD Collector, Ceramic Insulators,
and Jet
When replacing the jet, always install a new collector, ceramic insulators, and metal C-rings.
When replacing the collector, Agilent recommends replacing the ceramic insulators and metal
C-rings.
The insulation around the inlets, detectors, valve box, and the insulation cups is made of
WAR N IN G
refractory ceramic fibers. To avoid inhaling fiber particles, we recommend the following
safety procedures: ventilate your work area; wear long sleeves, gloves, safety glasses, and a
disposable dust/mist respirator; dispose of insulation in a sealed plastic bag; wash your
hands with mild soap and cold water after handling the insulation.
1 Gather the following:
The bead is delicate. Be careful not to break or crack the bead. When performing
C AU T I O N
maintenance on the NPD, avoid touching the bead with your fingers, and prevent it from
coming in contact with other surfaces.
2 Turn off the bead and disable Adjust Offset.
3 Check and note the NPD leakage current for reference. See
4 Launch the GC maintenance wizard: Maintenance > Detector> Perform Maintenance >
196
NPD ceramic insulator kit, see
Collector
Cap for the bead
T-10 and T-20 Torx screwdrivers
Tweezers
Cotton swab
Solvent
Methanol
Jet
Lint-free gloves
Compressed, filtered dry air or nitrogen
Current"
on page 201.
Replace NPD Ceramics> Start Maintenance. Wait for the GC to become ready. The wizard
will guide you through the maintenance steps described below.
"Consumables and Parts for the NPD"
on page 186.
"To Check the NPD Leakage
Maintaining Your GC