Maintaining the Purged Packed Inlet

To Change the O-Ring on the Purged Packed Inlet

To Change the O-Ring on the Purged Packed Inlet

1 Gather the following:
2 Launch the GC maintenance wizard: Maintenance > Inlets, select the inlet, then select
Be careful! The oven and/or inlet may be hot enough to cause burns. If the inlet is hot, wear
heat-resistant gloves to protect your hands.
WAR N IN G
3 Loosen the top insert assembly to remove the top portion of the inlet.
Wear clean, lint-free gloves to prevent contamination of parts with dirt and skin oils.
C AU T I O N
4 Use tweezers to remove the old O-ring.
5 Insert a new O-ring.
6 Install and tighten the top insert assembly.
7 The GC maintenance wizard will perform checks at the appropriate times, including Leak &
8 Select Finished, then select OK to exit the GC maintenance wizard.
9 Restore the analytical method.
Maintaining Your GC
Replacement O-ring, see
page 66.
Septum nut wrench
Tweezers
Lint-free gloves
Perform Maintenance > Replace O-Ring > Start Maintenance. Wait for the GC to become
ready. The wizard will guide you through the maintenance steps described below.
Restriction tests, and will automatically reset the maintenance counters.
"Consumables and Parts for the Purged Packed Inlet"