To Change The Liner And O-Ring On The Split/Splitless Inlet - Agilent Technologies 8860 Maintaining

Gas chromatograph
Hide thumbs Also See for 8860:
Table of Contents

Advertisement

5
Maintaining the Split/Splitless Inlet

To Change the Liner and O-Ring on the Split/Splitless Inlet

To Change the Liner and O-Ring on the
Split/Splitless Inlet
1 Gather the following:
2 Launch the GC maintenance wizard: Maintenance > Inlets, select the inlet, then select
Be careful! The oven and/or inlet may be hot enough to cause burns. If either is hot, wear
heat-resistant gloves to protect your hands.
WAR N IN G
3 Slide the locking tab forward (counterclockwise). Lift the insert assembly straight up and
54
Replacement O-ring, see
page 42.
Replacement liner
Tweezers
Wrench, hex for changing septum (optional)
Wrench, capillary inlet (optional)
Lint-free gloves
Perform Maintenance > Replace Liner > Start Maintenance. The wizard will walk through
the steps needed to replace the liner and O-ring (those same steps are repeated below).
away from the inlet to avoid chipping or breaking the liner.
"Consumables and Parts for the Split/Splitless Inlet"
on
Maintaining Your GC