8
Maintaining the COC Inlet
To Install an Insert on the COC Inlet
7 Check that the insert is the correct size for both the needle and column. See
8 Lower the new insert straight into the inlet from the top. The insert can be installed either
9 Install the spring on top of the insert.
10 Install the septum and septum nut or duckbill septum and cooling tower assembly and
Execute the column installation procedure manually. Do not use the automated column
NOTE
installation wizard.
11 Install the column. See
12 Exit maintenance mode. Select Maintenance > Instrument > Perform Maintenance >
Maintaining Your GC
Needle-to-Column Size on the COC Inlet"
end up.
finger-tighten.
"To Install a Capillary Column with the Cool On-Column Inlet"
page 124.
Maintenance Mode > Finished.
on page 127.
"To Check the
on