Agilent Technologies 8860 Maintaining page 91

Gas chromatograph
Hide thumbs Also See for 8860:
Table of Contents

Advertisement

6
Maintaining the Purged Packed Inlet
To Install a Packed Glass Column
8 Withdraw the column 1 to 2 mm from both the inlet and detector. Finger-tighten both
Overtightening the column nut or forcing it to bottom in either the inlet or detector may
C AU T I O N
shatter the column.
9 Tighten both column nuts 1/4 turn with a wrench.
10 The GC maintenance wizard will perform checks at the appropriate times, including Leak &
11 Select Finished, then select OK to exit the GC maintenance wizard.
12 Configure the new packed column (make sure either column length or diameter is zero).
13 Establish a flow of carrier gas and purge as recommended by the packing manufacturer.
14 Restore the analytical method.
15 After the GC becomes ready, wait 10 minutes then ignite the detector flame or adjust
Be careful! The oven, inlet, and/or detector may be hot enough to cause burns. If the oven,
WAR N IN G
inlet, or detector is hot, wear heat-resistant gloves to protect your hands.
16 Allow the oven, inlet, and detector to equilibrate at operating temperature, then retighten
17 Perform the inlet Leak & Restriction test and reset the maintenance counters.
Maintaining Your GC
column nuts.
Restriction tests, and will automatically reset the maintenance counters.
Generally:
20 to 30 mL/min for 2-mm id glass or 1/8-inch od metal columns
50 to 60 mL/min for 4-mm id glass or 1/4-inch od metal columns
For FPD+, immediately turn off the flame.
For NPD, immediately turn off the bead.
offset on the NPD bead.
the fittings.
91

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents