Agilent Technologies 8860 Maintaining page 133

Gas chromatograph
Hide thumbs Also See for 8860:
Table of Contents

Advertisement

8
Maintaining the COC Inlet
To Clean the COC Inlet
8 Remove the spring and insert from the bath and rinse thoroughly with water and methanol.
9 Dry the spring and insert with compressed air or nitrogen.
10 Install the insert. See
11 Install the column. See
12 Exit maintenance mode. Select Maintenance > Instrument > Perform Maintenance >
Maintaining Your GC
"To Install an Insert on the COC Inlet"
"To Install a Capillary Column with the Cool On-Column Inlet"
page 124.
Maintenance Mode > Finished.
on page 130.
on
133