Maintaining the Purged Packed Inlet

To Change the Glass Liner on the Purged Packed Inlet

To Change the Glass Liner on the Purged Packed
Inlet
1 Gather the following:
2 Launch the GC maintenance wizard: Maintenance > Inlets, select the inlet, then select
Be careful! The oven and/or inlet may be hot enough to cause burns. If the inlet is hot, wear
heat-resistant gloves to protect your hands.
WAR N IN G
3 Loosen the top insert assembly to remove the top portion of the inlet.
Top insert weldment
Flared end
Glass liner
4 Use a thin wire or wood splint to carefully lift and remove the old glass liner.
80
Replacement glass liner, see
page 66.
9/16-inch wrench
Lint-free gloves
Perform Maintenance > Replace Liner > Start Maintenance. Wait for the GC to become
ready. The wizard will guide you through the maintenance steps described below.
"Consumables and Parts for the Purged Packed Inlet"