To Cut A Loop From A Column - Agilent Technologies 8860 Maintaining

Gas chromatograph
Hide thumbs Also See for 8860:
Table of Contents

Advertisement

4
Maintaining Capillary Columns

To Cut a Loop from a Column

To Cut a Loop from a Column

1 Gather the following:
2 Place GC in maintenance mode: Maintenance > Instrument > Perform Maintenance >
Be careful! The oven and/or inlet may be hot enough to cause burns. If the inlet is hot, wear
WAR N IN G
heat-resistant gloves to protect your hands.
Wear safety glasses to protect your eyes from flying particles while handling, cutting, or
WAR N IN G
installing glass or fused silica capillary columns. Use care in handling these columns to
prevent puncture wounds.
3 Loosen the inlet column nut and remove the column from the inlet.
4 Uncoil one loop of column from the column hanger.
5 Cut the unwanted loop from the column.
6 Install the column into the inlet using the new ferrules.
Perform the installation procedure manually. Do not use the automated column installation
NOTE
wizard.
7 Exit maintenance mode. Select Maintenance > Instrument > Perform Maintenance >
Maintaining Your GC
New ferrule(s) for the column inlet connection
Column cutter
Maintenance Mode > Start Maintenance and wait for the GC to become ready.
"To Install a Capillary Column with the Split/Splitless Inlet"
"To Install a Capillary Column with the Purged Packed Inlet"
"To Install a Capillary Column with the Cool On-Column Inlet"
Maintenance Mode > Finished.
37

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents