Agilent Technologies 8860 Maintaining page 241

Gas chromatograph
Hide thumbs Also See for 8860:
Table of Contents

Advertisement

A
Swagelok Connections
Making Swagelok Connections
5 Push the tube fully into the plug, then withdraw it approximately 1 to 2 mm as shown in
Front
ferrule
Figure 25.
6 Finger-tighten the nut.
7 Mark the nut with a pencil line. See
Pencil line
Figure 26.
8 For 1/8-inch Swagelok fittings, use a pair of 7/16-inch wrenches to tighten the fitting 3/4 of
Figure 27.
Maintaining Your GC
Figure
25.
Insert the tubing
Mark the fitting
a turn. See
Figure
27.
For 1/4-inch fittings, use a pair of 9/16-inch wrenches to tighten them 1-1/4 turn, as
shown in
Figure
27.
Final tightening
Nut
Insert tubing fully
Back
ferrule
Withdraw tubing 1 to 2 mm
Tighten nut
Figure
26.