5
Maintaining the Split/Splitless Inlet
To Replace the Gold Seal on the Split/Splitless Inlet
Wear clean, lint-free gloves to prevent contamination of parts with dirt and skin oils.
C AU T I O N
6 Put on gloves to protect the new gold seal and washer from contamination. Put a new
Inlet gold seal
Washer
Reducing nut
7 Replace the reducing nut and tighten securely with a wrench.
8 Replace the inlet liner.
9 Install the column and the insulation cup.
10 Bakeout contaminants. See
11 Perform an inlet Leak & Restriction test and reset the maintenance counters.
12 Select Finished, then select OK to exit the GC installation wizard.
13 Restore the analytical method.
58
washer in the reducing nut and place the new gold seal on top of it (raised portion facing
down).
page 63.
Gold seal side view
Make sure raised portion faces
down.
"To Bakeout Contaminants from the Split/Splitless Inlet"
on
Maintaining Your GC