7
Maintaining the Packed Column Inlet
To Change the Septum on the Packed Column Inlet
6 Replace the septum retainer nut or Merlin cap and finger-tighten. Tighten the septum
Overtightening the septum nut can cause contamination.
C AU T I O N
7 Reset the septum counter.
8 The GC maintenance wizard will perform checks at the appropriate times, including Leak &
9 Select Finished, then select OK to exit the GC maintenance wizard.
10 Restore the analytical method.
100
retainer nut until the C-ring is about 1 mm above the nut.
1 mm
Restriction tests, and will automatically reset the maintenance counters.
Maintaining Your GC