Maintaining the Split/Splitless Inlet
To Clean the Split/Splitless Inlet
13 Exit maintenance mode. Select Maintenance > Instrument > Perform Maintenance >
14 Bakeout contaminants. See
15 Perform an inlet Leak & Restriction test.
16 On the GC, select Maintenance > Inlets, then reset the counters for the parts you changed.
17 Restore the analytical method.
62
Maintenance Mode > Finished.
page 63.
"To Bakeout Contaminants from the Split/Splitless Inlet"
on
Maintaining Your GC