Agilent Technologies 8860 Maintaining page 149

Gas chromatograph
Hide thumbs Also See for 8860:
Table of Contents

Advertisement

9
Maintaining the FID
To Replace an FID Jet
10 Install the castle assembly.
11 The GC maintenance wizard will perform checks at the appropriate times, including Leak &
12 Select Finished, then select OK to exit the GC maintenance wizard.
13 Attach the capillary column to the detector.
Be careful! The oven, inlet, and/or detector may be hot enough to cause burns. If the oven,
WAR N IN G
inlet, or detector is hot, wear heat-resistant gloves to protect your hands.
14 Check the FID baseline. See
Maintaining Your GC
Restriction tests, and will automatically reset the maintenance counters.
a Install the column in the detector. See
page 144.
b After the column is installed at both inlet and detector, establish a flow of carrier gas
and purge as recommended by the column manufacturer.
c Check the FID leakage current. See
d Bakeout the detector. See
e Restore the analytical method.
f Allow the oven, inlet, and detector to equilibrate at operating temperature, then
retighten the fittings.
"To Install a Capillary Column in the FID"
"To Check the FID Leakage Current"
"To Bakeout the FID"
on page 161.
"To Check the FID Baseline"
on page 158.
on page 159.
on
149