13 Maintaining the FPD+
To Change the FPD+ Wavelength Filter
8 Install the filter in the filter housing. Install the sulfur filter spacer, if necessary.
9 Replace the PMT assembly and secure with the spring.
10 Route the PMT wires through the clips as shown. Avoid placing the wires very near heated
11 The GC maintenance wizard will perform checks at the appropriate times, including Leak &
12 Select Finished, then select OK to exit the GC maintenance wizard.
13 Restore the analytical method.
Maintaining Your GC
areas (such as the emission block or oven top).
Restriction tests, and will automatically reset the maintenance counters.
211