9
Maintaining the FID
To Perform Maintenance on the FID Collector Assembly
16 Install the FID ignitor.
Maintaining Your GC
b Insert the FID castle into the collector housing.
c Install the spring washer over the castle.
d Install the collector nut over the castle and tighten firmly. The seal should be airtight.
Maintain the orientation of the ignitor hole with the base as shown below.
a Insert the ignitor and copper seal into the threaded hole of the castle. Keep the mating
threads clean.