Agilent Technologies 8860 Maintaining page 126

Gas chromatograph
Hide thumbs Also See for 8860:
Table of Contents

Advertisement

8
Maintaining the COC Inlet
To Install a Capillary Column with the Cool On-Column Inlet
12 Configure the new column.
13 Condition the column per the manufacturer's recommendation. See
14 Install the column into the detector.
Execute the column installation procedure manually. Do not use the automated column
NOTE
installation wizard.
15 The GC maintenance wizard will perform checks at the appropriate times, including Leak &
16 Select Finished, then select OK to exit the GC maintenance wizard.
17 Restore the analytical method.
18 After the column is installed at both inlet and detector, establish a flow of carrier gas and
19 Restore the analytical method.
20 After the GC becomes ready, wait 10 minutes then ignite the detector flame or adjust
Be careful! The oven, inlet, and/or detector may be hot enough to cause burns. If they are hot,
WAR N IN G
wear heat-resistant gloves to protect your hands.
21 Allow the oven, inlet, and detector to equilibrate at operating temperature, then retighten
126
Capillary
Column.
To Install a Capillary Column in the FID
To Install a Capillary Column in the NPD
To Install a Capillary Column in the TCD
To Install a Capillary Column in the ECD
To Install a Capillary Column to the FPD+
Restriction tests, and will automatically reset the maintenance counters.
purge as recommended by the column manufacturer.
For FPD+, immediately turn off the flame.
For NPD, immediately turn off the bead
offset on the NPD bead.
the fittings.
To Condition a
Maintaining Your GC