14 Maintaining EPC Modules
Installing or Replacing Frits in the PCM
When hydrogen is used, dangerously high flows are possible if insufficient flow resistance is
WAR N IN G
provided downstream of the supply tube. Always use either the High (Blue dot) or Medium
(Red dot) frit with hydrogen.
After installing or replacing a frit, be sure to update the PIDs used with your PCM.
Maintaining Your GC
225