10 Maintaining the TCD

To Install a Capillary Column in the TCD

To Install a Capillary Column in the TCD
1 Gather the following materials, see
2 Launch the GC maintenance wizard: Maintenance > Column > Perform Maintenance >
Be careful! The oven, inlet, and/or detector may be hot enough to cause burns. If the oven,
WAR N IN G
inlet, or detector is hot, wear heat-resistant gloves to protect your hands.
Wear safety glasses to protect your eyes from flying particles while handling, cutting, or
WAR N IN G
installing glass or fused silica capillary columns. Use care in handling these columns to
prevent puncture wounds.
Wear clean, lint-free gloves to prevent contamination of parts with dirt and skin oils.
C AU T I O N
3 Place a septum (if the column id is ≤0.1 mm), capillary column nut, and ferrule on the
Ferrule
Column nut
Septum
4 Score the column using a glass scribing tool. The score must be square to ensure a clean
166
Detector"
on page 164:
Column
Ferrule(s)
Column nut
Column cutter
1/4-inch open-end wrench
Septum
Isopropanol
Lab tissue
Lint-free gloves
Install Column > Start Maintenance. Wait for the GC to become ready.
The wizard will guide you through the maintenance steps described below.
column.
break.
"Consumables and Parts for the Thermal Conductivity