11 Maintaining the ECD

To Replace the ECD Fused Silica Indented Mixing Liner and Install the Makeup Gas Adapter

To Replace the ECD Fused Silica Indented Mixing
Liner and Install the Makeup Gas Adapter
1 Gather the following:
2 Place GC in maintenance mode: Maintenance > Instrument > Perform Maintenance >
Be careful! The oven and/or detector may be hot enough to cause burns. If the detector is
WAR N IN G
hot, wear gloves to protect your hands.
Wear safety glasses to protect your eyes from flying particles while handling, cutting, or
WAR N IN G
installing glass or fused silica capillary columns. Use care in handling these columns to
prevent puncture wounds.
3 Remove the column from the makeup gas adapter.
To prevent damage avoid flexing/bending the tubing on the makeup gas adapter.
C AU T I O N
4 Remove the makeup gas adapter.
Maintaining Your GC
Fused silica indented mixing liner, see
page 174.
1/4-inch Swagelok nut
1/4-inch Vespel/graphite ferrule
9/16-inch wrench
Methanol
Lint-free gloves
Maintenance Mode > Start Maintenance. Wait for the GC to become ready to continue (all
zones cooled to safe handling temperatures).
a Loosen the adapter nut with a wrench and slide out the makeup gas adapter from the
ECD. Remove the ferrule.
The makeup gas adapter will remain attached to the supply tubing and hang
suspended in the oven.
b Adjust the adapter's position so that maintenance can be performed on the adapter
easily and without obstruction.
"Consumables and Parts for the ECD"
on
177