13 Maintaining the FPD+
To Replace the FPD+ Ignitor
6 Install the new ignitor assembly and tighten with the nut driver. Do not overtighten. (If the
7 Replace the ignitor collet and cable assembly and tighten the screw. Align the collet set
8 Replace the FPD cover. See
9 The GC maintenance wizard will perform checks at the appropriate times, including Leak &
10 Select Finished, then select OK to exit the GC maintenance wizard.
11 Restore the analytical method.
12 Wait 20 min. for the detector to heat up, then ignite the flame.
214
ignitor comes with a copper washer, discard the copper washer.)
screw as shown. Do not let the collet screw touch metal parts, such as the emission block
or PMT bracket (dual wavelength detector).
Restriction tests, and will automatically reset the maintenance counters.
"To Install the FPD+ Cover"
on page 215.