Maintaining the Packed Column Inlet

To Install a Packed Metal Column

To Install a Packed Metal Column

1 Gather the following:
2 Place GC in maintenance mode: Maintenance > Instrument > Perform Maintenance >
3 Prepare the packed metal column. See
Be careful! The oven, inlet, and/or detector may be hot enough to cause burns. If the oven,
WAR N IN G
inlet, or detector is hot, wear heat-resistant gloves to protect your hands.
4 Install the 1/8-inch or 1/4-inch packed column inlet adapter, if necessary. See
Wear clean, lint-free gloves to prevent contamination of parts with dirt and skin oils.
C AU T I O N
5 Attach the column to the inlet adapter. Finger-tighten the nut.
6 Tighten the nut an additional 1/4 turn with a wrench (for a 1/8-inch column) or an
7 Configure the new packed column (make sure either column length or diameter is zero).
8 Exit maintenance mode. Select Maintenance > Instrument > Perform Maintenance >
Do not use hydrogen as the carrier for conditioning! It could vent into the oven and present an
WAR N IN G
explosion hazard.
9 Condition the column, if necessary. See
10 Place GC in maintenance mode: Maintenance > Instrument > Perform Maintenance >
Be careful! The oven, inlet, and/or detector may be hot enough to cause burns. If the oven,
WAR N IN G
inlet, or detector is hot, wear heat-resistant gloves to protect your hands.
11 If required, install a packed column adapter onto the detector column fitting. See
12 Place a nut and ferrule onto the packed column.
13 Attach the column to the detector or detector adapter. Finger-tighten the nut.
Maintaining Your GC
7/16-inch, 9/16-inch, and 1/2-inch wrenches
Lint-free gloves
Maintenance Mode > Start Maintenance. Wait for the GC to become ready.
page 118.
an Adapter on the Packed Column Inlet"
additional 3/4 turn (for a 1/4-inch column).
Use two wrenches, one on the column nut and the other on the adapter, to prevent the
adapter from rotating.
Maintenance Mode > Finished.
Maintenance Mode > Start Maintenance. Wait for the GC to become ready.
Install a Packed Column Adapter on a Detector Fitting"
"To Install Ferrules on a Packed Metal Column"
on page 103.
"To Condition a Packed Column"
on page 113.
on
"To Install
on page 117.
"To
111