To Install Ferrules On A Packed Metal Column - Agilent Technologies 8860 Maintaining

Gas chromatograph
Hide thumbs Also See for 8860:
Table of Contents

Advertisement

6
Maintaining the Purged Packed Inlet

To Install Ferrules on a Packed Metal Column

To Install Ferrules on a Packed Metal Column

1 Gather the following:
2 Verify that the column end is cut square and is free of burns and deformation.
3 Secure the fitting in a bench vise.
Wear clean, lint-free gloves to prevent contamination of parts with dirt and skin oils.
C AU T I O N
4 Assemble a Swagelok nut and ferrules onto the column.
Front ferrule
Back ferrule
Column nut
5 Fully insert the column into the vise-held fitting, then withdraw 1–2 mm. Finger-tighten the
6 Tighten the nut an additional 3/4 turn with a wrench (for a 1/8-inch column) or an
7 Unscrew the column nut from the vise-held fitting and remove the column. Ferrules should
Maintaining Your GC
Wrenches
Stainless steel male Swagelok fitting, 1/4- or 1/8-inch od
Brass Swagelok nut and ferrule set, see
Packed Inlet"
on page 66.
Lint-free gloves
nut.
additional 1-1/4 turn (for a 1/4-inch column).
now be set in place on the column with the column end correctly positioned.
"Consumables and Parts for the Purged
93