To Remove The Fpd Cover; To Remove The Fpd + Cover - Agilent Technologies Intuvo 9000 Series Maintaining

Gas chromatograph
Table of Contents

Advertisement

To Remove the FPD
C A U T I O N
WA RNING
Maintaining Your GC
+
Cover
1
Gather the following:
T-20 Torx screwdriver
2
Prepare for maintenance. On the GC touch display, go to
Maintenance > Instrument > Perform Maintenance > Maintenance
Mode> Start Maintenance. Wait for the GC to become ready.
When turning off the GC, first turn off the flame to prevent
condensation from dripping into the jet and column.
Be careful! The detector may be hot enough to cause burns. If the
detector is hot, wear heat-resistant gloves to protect your hands.
3
Open the FPD detector top cover.
4
Loosen the screws securing the FPD cover to the top of the
detector.
+
Maintaining the FPD
151