9
Maintaining the FID
To Perform Maintenance on the FID Collector Assembly
4 If replacing only the FID ignitor assembly with copper washer, skip to
5 Remove the three screws that secure the collector assembly to the FID mounting pallet.
This step exposes the interconnect spring. Be careful not to touch or disfigure the spring
C AU T I O N
while working on the FID. Any dirt or bending will reduce the sensitivity of your detector.
6 Remove the collector assembly. Place it on a clean cloth for additional disassembly.
Maintaining Your GC
b Loosen the ignitor with a wrench.
c Turn the nut counterclockwise by hand. Remove the ignitor and copper washer.
step 16
for assembly.