Maintaining the Packed Column Inlet
To Change the Glass Liner on the Packed Column Inlet
6 Wearing gloves, grasp the flared end (top) of the replacement glass liner with tweezers and
7 Install the top insert weldment and hand-tighten firmly.
8 Exit maintenance mode. Select Maintenance > Instrument > Perform Maintenance >
9 Reset the EMF counter.
10 Perform an inlet Leak & Restriction test and reset the maintenance counters.
11 Restore the analytical method.
106
install it in the inlet.
Maintenance Mode > Finished.