9
Maintaining the FID
To Perform Maintenance on the FID Collector Assembly
14 Install the collector insulators.
15 Install the FID ignitor castle.
154
a Insert one of the insulators into the base of the housing. Seat the insulator with the flat
surface facing out of the housing.
b Insert the long end of collector into the housing and lower insulator.
c Insert the other insulator onto the top of the collector, with the flat surface facing
towards the housing.
a Orient the castle so that the threaded hole for the ignitor faces toward the electronics.