9
Maintaining the FID
To Replace an FID Jet
To Replace an FID Jet
1 Gather the following:
2 Launch the GC maintenance wizard: Maintenance > Detectors, select the detector, then
Be careful! The oven and/or detector may be hot enough to cause burns. If the detector is
hot, wear gloves to protect your hands.
WAR N IN G
Wear safety glasses to protect your eyes from flying particles while handling, cutting, or
WAR N IN G
installing glass or fused silica capillary columns. Use care in handling these columns to
prevent puncture wounds.
3 If installed, remove the capillary column from the detector.
Wear clean, lint-free gloves to prevent contamination of parts with dirt and skin oils.
C AU T I O N
4 Remove the FID castle assembly and place it on a clean cloth.
5 Locate the jet inside the housing.
Maintaining Your GC
•
Replacement jet
•
T-20 Torx screwdriver
•
1/4-inch nut driver
•
Tweezers
•
Compressed, filtered, dry air or nitrogen
•
Solvent that will clean the type of deposits in your detector
•
Clean cloth
•
Cotton swab
•
Lint-free gloves
select Perform Maintenance > Replace FID Jet > Start Maintenance. The wizard will guide
you through the maintenance steps described below.
147