6
Maintaining the Purged Packed Inlet

To Install an Adapter on the Purged Packed Inlet

To Install an Adapter on the Purged Packed Inlet

1 Gather the following:
2 Place GC in maintenance mode: Maintenance > Instrument > Perform Maintenance >
Be careful! The oven and/or inlet may be hot enough to cause burns. If the inlet is hot, wear
WAR N IN G
heat-resistant gloves to protect your hands.
Wear clean, lint-free gloves to prevent contamination of parts with dirt and skin oils.
C AU T I O N
3 Clean the end of the adapter with a lint-free cloth and methanol to remove contamination
4 Place the tubing nut and Vespel/graphite ferrule on the adapter.
Inlet fitting
Ferrule
Tubing nut
Adapter
5 Insert the adapter straight into the inlet base as far as possible.
6 Hold the adapter in this position and finger-tighten the nut.
7 Tighten an additional 1/4 turn with a wrench.
8 Exit maintenance mode. Select Maintenance > Instrument > Perform Maintenance >
78
Brass tubing nut, see
page 66.
Adapter (0.53 mm, 1/8-inch packed, or 1/4-inch packed)
7/16-inch and 9/16-inch wrench
Vespel/graphite ferrule
Methanol
Lint-free gloves
Maintenance Mode > Start Maintenance. Wait for the GC to become ready.
such as fingerprints.
Maintenance Mode > Finished.
"Consumables and Parts for the Purged Packed Inlet"
on
Maintaining Your GC