To Replace The 7693A Injector Needle Support Guide - Agilent Technologies 8860 Maintaining

Gas chromatograph
Hide thumbs Also See for 8860:
Table of Contents

Advertisement

8
Maintaining the COC Inlet

To Replace the 7693A Injector Needle Support Guide

To Replace the 7693A Injector Needle Support
Guide
Before using a 7693A ALS to make injections onto the COC inlet, install the on-column needle
support guide.
Refer to the 7693A ALS documentation for complete details.
1 Open the injector door.
2 Remove the syringe.
3 Slide the syringe carriage up to the top position.
4 Completely remove the T-10 Torx screw from the support foot. Be careful to not let the
5 Slide off the support foot.
6 Slide on the new support foot.
7 Replace the T-10 Torx screw and tighten.
8 Install the appropriate syringe.
9 Close the injector door.
10 Align the injector.
134
screw fall into the turret assembly.
Standard needle
support guide
On-column
needle support
guide
Maintaining Your GC