To Bakeout Contaminants From The Packed Column Inlet - Agilent Technologies 8860 Maintaining

Gas chromatograph
Hide thumbs Also See for 8860:
Table of Contents

Advertisement

7
Maintaining the Packed Column Inlet

To Bakeout Contaminants from the Packed Column Inlet

To Bakeout Contaminants from the Packed
Column Inlet
1 Set the column flow to the normal operating value.
2 Purge the column with carrier flow for at least 10 minutes before heating the oven.
3 If the column is attached to the detector, set the detector 25 °C above normal operating
Be careful! The oven, inlet, and/or detector may be hot enough to cause burns. If they are hot,
WAR N IN G
wear heat-resistant gloves to protect your hands.
4 Set the inlet temperature to 300 °C or 25 °C above the normal operating temperature.
5 Set the column oven 25 °C above the GC method final oven temperature to bake
6 Bakeout for 30 minutes or until the detector baseline is free of contamination peaks.
110
temperature.
If the column is not attached to the detector, cap the detector fitting.
contaminants out of the inlet. Do not exceed the column manufacturer's maximum
temperature limit.
Maintaining Your GC