Agilent Technologies 8860 Maintaining page 89

Gas chromatograph
Hide thumbs Also See for 8860:
Table of Contents

Advertisement

6
Maintaining the Purged Packed Inlet
To Install a Packed Column Adapter onto a Detector
Detector fitting
5 Exit maintenance mode. Select Maintenance > Instrument > Perform Maintenance >
After installation, heat the oven to 200 °C, cool it to a safe handling temperature, then re-tighten
the column connection. (Re-tighten until snug, only about 15 degrees rotation.) Repeat two
more times. This process will set the Vespel/graphite ferrule to help prevent leaks.
Maintaining Your GC
Maintenance Mode > Finished.
89