Agilent Technologies 8860 Maintaining page 156

Gas chromatograph
Hide thumbs Also See for 8860:
Table of Contents

Advertisement

9
Maintaining the FID
To Perform Maintenance on the FID Collector Assembly
17 Lower the collector assembly into the housing.
18 Insert the three screws and tighten (to 18-inch-pounds).
19 Connect the ignitor extension cable.
20 Verify assembly:
156
b Tighten the ignitor with a wrench. Ignition requires a good electrical contact that is
free of any dirt.
a Check the FID leakage current. See
"To Check the FID Leakage Current"
on page 158.
Maintaining Your GC

Hide quick links: