9
Maintaining the FID
To Perform Maintenance on the FID Collector Assembly
7 Remove the gasket from the bottom of the assembly, if necessary.
8 Remove the FID ignitor castle.
9 If only replacing the FID castle, skip to
152
a Loosen the collector nut.
b Remove the collector nut and the spring washer.
c Lift the castle out of the collector housing. When removing the castle, some of the
collector parts may be attached. Set these on a clean cloth to protect from scratches
or dirt.
step 15
for reassembly.