3
Troubleshooting and Test Functions
Shutdown
An external instrument has generated a shut-down signal on the remote line.
The sampler continually monitors the remote input connectors for status
signals. A LOW signal input on pin 4 of the remote connector generates the
error message.
Probable Causes
• Leak detected in another Agilent 1100 module with a CAN connection to
• Error detected in an external instrument with a remote connection to the
• The degasser failed to generate sufficient vacuum for solvent degassing.
Suggested Actions
62
the system.
system.
Determine which Agilent 1100 module has the leak. Fix the leak before
restarting the well-plate sampler.
Check external instruments for an error condition.
Check the degasser for an error condition. Refer to the Reference Manual
for the Agilent 1100 Series degasser.
1100 Series WS MWS Reference Manual