Installing the Sampler
14 Connect the power cables to the power connectors.
15 Connect the CAN cable to other Agilent 1100 series modules.
16 If a Agilent ChemStation is the controller, connect either
17 Connect the APG remote cable (optional) for non Agilent 1100 Series
18 Ensure the side panel is correctly installed.
19 Turn ON power by pushing the button at the lower left hand side of the
20 Close the front door. The exhaust fan will turn ON and remove the vapor
The sampler is turned ON when the line power switch is pressed and the green indicator
NO TE
lamp is illuminated. The detector is turned 0FF when the line power switch is protruding
and the green light is 0FF.
To disconnect the sampler from the line, unplug the power cord. The power supply
WA RN ING
still uses some power, even if the power switch at the front panel is turned 0FF.
To avoid personal injury, keep fingers away from the needle area during autosampler
WA RN ING
operation. Do not attempt to insert or remove a vial or a plate when the needle is
positioned.
28
the GPIB cable to the detector
the LAN connection to the LAN interface
instruments.
sampler.
from the tray compartment. After 1-2 minutes the sampler will start tile
hardware initialisation process. At the end of this process the status LED
should be green.
1100 Series WS MWS Reference Manual