Fan Failed - Agilent Technologies G1367A Reference Manual

1100 series well-plate sampler & micro well-plate sampler
Table of Contents

Advertisement

3
Troubleshooting and Test Functions

Fan Failed

The cooling fan in the sampler has failed.
The hall sensor on the fan shaft is used by the MTP board to monitor the fan
speed. If the fan speed falls below 2 revolutions/second for longer than
5 seconds, the error message is generated.
Probable Causes
• Fan cable disconnected.
• Defective fan.
• Defective MTP board.
Suggested Actions
70
Ensure the fan is connected correctly.
Exchange fan.
Exchange the MTP board.