Sample Transport Self Alignment

The sample transport alignment with the sampling unit and the well-plate tray
is required to compensate for larger deviations in positioning the needle
carrier.This function is in the diagnose screen of the Chemstation or the
Control Module.
The sample transport self alignment is required after disassembling the
system or when you exchange:
• The sample transport.
• The sampling unit.
• The MTP main board.
• The well plate sampler tray
The sample transport self alignment requires the standard well-plate tray (Part
WA RN ING
Number: G1367-60001)
The sample transport self alignment must be done with an empty WPS tray
WA RN ING
1100 Series WS MWS Reference Manual
Arm position in X
Arm position in
Theta
Left side
Straight
Left side
Straight
Middle
Left
Right side
Left rear
Left side
Right rear
Troubleshooting and Test Functions
Arm Position in Z
Note
Up
No current on Theta
Middle
No current on the ST
Up
Up
Up